Home » Product List » Multi-Channels Syringe Pump » China Manufacturer of Syringe Pump
Product Details
China Manufacturer of Syringe Pump
| Standard: | CE Certificate |
|---|---|
| Productivity: | 1000 Sets/Month |
| Trademark: | SK |
| Origin: | China |
| Export Markets: | North America, South America, Eastern Europe, Southeast Asia, Africa, Oceania, Mid East, Eastern Asia, Western Europe |
SK-500III Double Channels Syringe Pump
Features and Functions:
1. Multi-channels(Double-channels), sub-rate injection.
2. Four kinds of injection modes: Rate mode, Time mode, Body Weight mode, link mode.
3. Innovative design of outer appearance gives a strong sense of science and technology.
4. Double CPU makes the process of injection safer and more reliable.
5. Brand-new STM32 ARM-Cortex M3 platform, optimized circuit design, operation rate more faster and stable.
6. The use of TFT color LCD makes the screen display more remarkable.
7. Digital menu, easy operation; Functions fully meet the clinical demand.
8. Ample and user-friendly indicator information; Accurate and complete alarm information to ensure safe injection.
9 Aluminum alloy base, integrative pump device makes more smooth operation and easier installation and maintenance.
| Product type | SK-500III |
| Maximum injection rate | 1500ml/h (Different injectors have different maximum rate.) |
| Range of flow rate |
5ml injector: 0.1-100ml/h 10ml injector: 0.1-200ml/h 20ml injector: 0.1-400ml/h 30ml injector: 0.1-600ml/h 50ml injector: 0.1-1500ml/h |
| Injection increment | 0.1ml/h |
| Bolus rate |
5ml injector: 100ml/h 10ml injector: 200ml/h 20ml injector: 400ml/h 30ml injector: 400-600ml/h 50ml injector: 400-1500ml/h |
| Injection mode selection | 1. Rate mode 2. Time mode 3. Body Weight mode |
| Rate mode | Speed: 0.1-1500ml/h (maximum value depends on the injector specifications) |
| Time mode |
Time:1-2000 minutes Accumulated injection volume: 0.1-999.9ml |
| Body Weight mode |
Weight: 0.1-300.0kg Drug: 0.1-999.9(mg) Volume: 0.1-999.9ml Dose: 0.1-9999.9(restricted to unit, drug, solution, weight and injector specification). Unit: mg/kg/h, ug/kg/min |
| KVO rate | Adjustable range: 0.1ml/h-5.0ml/h (begin the KVO function under the occlusion situation) ; pressing STOP key to cancel the KVO function when it is unnecessary in use. |
| IP grade | IP21 |
| Injection pressure | The maximum pressure is 0.3Mpa, Occlusion alarm pressure value is 40kPa-160kPa when pressure value is minimum, injection rate is 25ml/h, the longest alarm time is 10 minutes. When pressure value is minimum, injection rate is 5ml/h, the longest alarm time is 10 minutes. |
| Preset volume | 0.1ml-9999.9ml |
| Preset time | 0-99 hours, 0-59 minutes |
| Accumulated injection volume | 0.1ml-9999.9ml |
| Accuracy | 3% |
| Power supply | AC 100-220V,50/60Hz |
| Battery | Rechargeable lithium polymer battery, 7.4 V, 1600mAh. |
| Maximum power consumption | 25VA, running more than 2 hours at the rate of 25ml/h after being fully charged. |
| Battery charge | When syringe pump is connected to AC power, the battery will automatically recharge. (about 8-14 hours to recharge fully.) |
| Fuses | T2AL250V- |
| Displayed information | Flow rate; accumulated injection volume; injector specification; battery capacity; bed No.; AC power indicator, operation indication, etc. |
| State indication information | Stop, inject, bolus, KVO (stop flashing, the other indicators flash by turns.) |
| Alarm information | near completion, Over, occlusion, low battery, abnormal 1 (communications failure), abnormal 2 (pump stuck), abnormal 3 (wrong parameter), volume limited alarm, no AC power |
| Maximum size of the outer shell | 288×130×122 mm(length×width×height) |
| Maximum weight | <3.0kg |
| Classification | ClassII, Type BF |
| Shell material | ABS plastic |
| Operating conditions | Environment temperature 5-40, air pressure 80kPa -106kPa,relative humidity ≤80% |
| Storage conditions | Environment temperature -40-55, air pressure 50kPa -106kPa,relative humidity ≤95% |
| Applicable syringes | 5ml.10 ml.20 ml.30ml.50 ml syringes |
| Electrical safety | In accordance with the stipulation of GB9706.1-1995 |
| Applied standards | According to the stipulation of YZB/Yue 001-2008 |
Inquiry Basket (